Summary
Background
Errors derive from the variability seen in the repeat volume measurements for a particular
individual. The aim of this study was to evaluate the variation and errors of a three-dimensional
(3D) multiple stereo camera system for objective breast assessment. It was also investigated
whether there was any correlation between these errors and the size of the breast.
Methods
A prototype eight camera multiple stereophotogrammetry system was utilized. The volumes
of nine plaster breast models were determined by 3D imaging and the correlation between
the size of the models and the variability of the measurements was investigated after
ten repeats.
The breasts of six live volunteers were examined following a specific protocol. The
breasts were captured six times, three times each on two different occasions; from
each breast capture a three-dimensional model was built and the breast volume was
measured three times with breast analysis tool (BAT) software. This allowed for an
assessment of the variability introduced at each stage of the measurement procedure.
The correlation between the size of the breast and the variability of the measurements
was investigated. Results of volume measurements by water displacement method, repeated
ten times with the plaster models and six times with the live models, were used for
comparison.
Results
The correlation between the size of the plaster models and the variability of the
measurements revealed a significant correlation (p = 0.033), indicating that the larger the model, the more variable were the results.
The correlation between the size of the breasts in the live models and the variability
of the results revealed a non-significant correlation (p = 0.342), but there was a visible trend that the larger breasts showed more variable
results.
The average variation in the repeated measurements on each individual was found to
be 32.95 cc across the two different poses, 19.43 cc across the various captures,
and 28.32 across the different volume calculations with the BAT software. The error
is less than what is clinically visible with the human eye, which is about 50 cc variation
in volume by subjective assessment. As a proportion of the mean volume, namely 6.9%,
4.0% and 5.9% respectably, the values are small and indicate that the reproducibility
of the system is good.
Conclusion
The 3D imaging system using multiple stereo cameras revealed a positive correlation
between the size of plaster models and the breasts of live models and the reproducibility
of the measurements, indicating that the variation was higher for the larger sizes.
The relationship was significant in the plaster models but not in the live models,
although a trend was observed.
The assessment of the variation and errors of the system was part of the necessary
procedure that should be considered for the validation of any new measurement technology
for breast assessment.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Plastic, Reconstructive & Aesthetic SurgeryAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Validation of a vision-based three dimensional facial imaging system.Cleft Palate Craniofac J. 2003; 40: 523-529
- Investigation into accuracy and reproducibility of a 3D Breast imaging system using multiple stereo cameras.JPRAS. 2011; 64: 577-582
- Validation and reproducibility of a high resolution three-dimensional facial imaging system.Br J Oral Maxillofac Surg. 2008; 46: 27-32
- Congenital breast asymmetry: subjective and objective assessment.Br J Plast Surg. 1994; 47: 95-102
Moffett RA. A prototype 3D breast surgery planning and assessment tool. Dissertation: University of Glasgow; October 1999.
- Breast volume determination in breast hypertrophy: an accurate method using two anthropomorphic measurements.Plast Reconstr Surg. 2006; 118: 313-320
Article info
Publication history
Published online: July 04, 2012
Accepted:
May 31,
2012
Received:
May 5,
2011
Identification
Copyright
© 2012 British Association of Plastic, Reconstructive and Aesthetic Surgeons. Published by Elsevier Inc. All rights reserved.