Research Article| Volume 56, ISSUE 7, P644-652, October 2003

Three-dimensional video-analysis of facial movements in healthy volunteers

      Abstract

      The aim of this study was to determine the mean distances of facial movements in 24 healthy individuals aged between 22 and 70 years, using the digitised three-dimensional video-analysis system developed by Frey et al. The subjects were divided into three groups of eight. The first group consisted of individuals aged between 20 and 30 years (mean±s.d.=25.0±2.33 years). Subjects in the second group were aged between 40 and 50 years (mean±s.d.=46.8±2.53 years), and the third group consisted of subjects aged between 60 and 70 years (mean±s.d.=63.6±3.07 years). In all groups the sexes were equally represented. No subject had had treatment to the face, nor did they have paralysis, scars or diseases of the skin. Males showed larger movements of the face than females, on average by 1.40±0.73 mm (15.08%). Subjects aged between 60 and 70 years demonstrated the largest movements of the face. The evaluation of facial movements in 24 healthy volunteers showed that sex and age affect facial dynamics. Thus study generated three-dimensional standard values for healthy facial movements.

      Keywords

      Three-dimensional (3D) quantitative analysis of movements is important in plastic and reconstructive facial surgery. In the past, several attempts have been made to measure the complex human face.
      • Takasaki H.
      Moire topography.
      • Burke P.H.
      • Banks P.
      • Beard L.F.
      • et al.
      Stereophotographic measurement of change in facial soft tissue morphology following surgery.
      • Deacon A.T.
      • Anthony A.G.
      • Bhatia S.N.
      • et al.
      Evaluation of a CCD-based facial measurement system.
      • Neely J.G.
      • Cheung J.Y.
      • Wood M.
      • et al.
      Computerized quantitative dynamic analysis of facial motion in the paralyzed and synkinetic face.
      • Sato K.
      • Inokuchi I.
      A range-imaging system utilizing a nematic liquid crystal mask.
      • Paletz J.L.
      • Manktelow R.T.
      • Chaban R.
      The shape of a normal smile: implications for facial paralysis reconstruction.
      • Bush K.
      • Antonyshyn O.
      Three-dimensional facial anthropometry using a laser surface scanner: validation of the technique.
      • Kohn L.A.
      • Cheverud J.M.
      • Bhatia G.
      • et al.
      Anthropometric optical surface imaging system repeatability, precision, and validation.
      • Ras F.
      • Habets L.L.
      • van Ginkel F.C.
      • et al.
      Quantification of facial morphology using stereophotogrammetry—demonstration of a new concept.
      • Bajaj-Luthra A.
      • Mueller T.
      • Johnson P.C.
      Quantitative analysis of facial motion components: anatomic and nonanatomic motion in normal persons and in patients with complete facial paralysis.
      • Bajaj-Luthra A.
      • Van Swearingen J.
      • Thornton R.H.
      • et al.
      Quantitation of patterns of facial movement in patients with ocular to oral synkinesis.
      • Meier-Gallati V.
      • Scriba H.
      • Fisch U.
      Objective scaling of facial nerve function based on area analysis (OSCAR).
      However, only a few have used instruments that provide 3D measurements and achieved promising results.
      • Caruso A.J.
      • Stanhope S.J.
      • McGuire D.A.
      A new technique for acquiring three-dimensional orofacial nonspeech movements.
      • Vannier M.W.
      • Pilgram T.K.
      • Bhatia G.
      • et al.
      Quantitative three-dimensional assessment of face-lift with an optical facial surface scanner.
      • Ferrario V.F.
      • Sforza C.
      • Poggio C.E.
      • et al.
      Distance from symmetry: a three-dimensional evaluation of facial asymmetry.
      • Trotman C.A.
      • Gross M.M.
      • Moffatt K.
      Reliability of a three-dimensional method for measuring facial animation: a case report.
      • Trotman C.A.
      • Faraway J.J.
      • Silvester K.T.
      • et al.
      Sensitivity of a method for the analysis of facial mobility. I. Vector of displacement.
      • Frey M.
      • Jenny A.
      • Giovanoli P.
      • et al.
      Development of a new documentation system for facial movements as a basis for the international registry for neuromuscular reconstruction in the face.
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      The aim of this study was to determine the mean standard values of healthy facial movements using the digitised 3D video-analysis system developed by Frey et al.
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      Throughout the text, values are given in the form mean±s.d.

      1. Material and methods

      We enrolled 24 subjects, aged between 22 and 70 years, in the study and divided them into three groups of eight. The first group consisted of individuals aged between 20 and 30 years (25.0±2.33 years). Subjects in the second group were aged between 40 and 50 years (46.8±2.53 years), and the third group consisted of subjects aged between 60 and 70 years (63.6±3.07 years). In all groups the sexes were equally represented. All 24 subjects were healthy and free from craniofacial anomalies. They had never undergone any treatment to the face nor had they had any form of facial paralysis, scarring or skin disease. Participants were of Middle European descent, had been born in Austria and lived nearby and in Vienna.
      The data-acquisition apparatus
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      consisted of a mirror system, a calibration grid (both developed by the Laboratory for Biomechanics, Swiss Federal Institute of Technology, Zurich, Switzerland) and a commercial digital video camera (Fig. 1) . We chose 18 standard reproducible anatomical landmarks on the face,
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      of which three were static and 15 were dynamic (Table 1 and Fig. 2) . A permanent marker was used to place a 2 mm black dot at each of the dynamic landmarks. Static points were marked with a plastic light ball 5 mm in diameter. All markings, except the points central nose and philtrum, were made on both sides of the face.
      Figure thumbnail gr1
      Fig. 1Diagram of the 3D video-analysis system (from Frey et al. (1999); reproduced with the permission of Lippincott Williams & Wilkins from Queen Arnette).
      Table 1Abbreviations of the standardised facial landmarks
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      Central noseCN
      Left ala of the noseLAN
      Left browLBP
      Left lower eyelidLLE
      Left mouth cornerLMC
      Left midlateral point of the lower lipLML
      Left midlateral point of the upper lipLMU
      Left tragusLTR
      Left upper eyelidLUE
      PhiltrumPH
      Right ala of the noseRAN
      Right browRBP
      Right lower eyelidRLE
      Right mouth cornerRMC
      Right midlateral point of the lower lipRML
      Right midlateral point of the upper lipRMU
      Right tragusRTR
      Right upper eyelidRUE
      Figure thumbnail gr2
      Fig. 2Locations of the dynamic and static landmarks on the face in frontal and lateral views.
      Each subject was videotaped under standard conditions. All recordings were made in the same room, in the same chair, by the same examiner and at the same time of day, between 14:00 and 16:00. Light from four 1000 W Halogen Photo-Optic lamps (Osram, Munich, Germany) was used to generate uniform, symmetrical and standardised lighting. Reflected light from the face was 5500 lux, measured by a hand-held photocell.
      The subject sat relaxed and upright in a normal chair without head support, eyes looking forward into the camera, which was positioned 5 m away. After the subject was positioned in the calibrated measurement field, she/he performed nine standard facial animations
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      (Table 2) in a sequential order after a verbal signal.
      Table 2Abbreviations of the standardised animations performed by all subjects in the order listed
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      English versionAbbreviationGerman version
      Maximal lifting of the eyebrowsmhaMaximales Heben der Augenbrauen
      Closure of the eyelids as in sleepsasSchliesen der Augen wie im Schlaf
      Maximal closure of the eyelidsmlMaximaler Lidschluss
      Maximal showing of the teethmzzMaximales Zähnezeigen
      Maximal showing of the teeth with maximal closure of the eyelidsmlmzGleichzeitig Maximaler Lidschluss und Maximales Zähnezeigen
      Smiling showing teethlzzLächeln mit Zeigen der Zähne
      Smiling with lips closedlglLächeln mit geschlossenen Lippen
      Pursing of the lipsmzpMund zuspitzen und pfeifen
      Pulling down the mouth cornersmwuMundwinkel nach unten ziehen
      Three repetitions of each set of facial animations were digitally collected and transferred to a computer, where the most suitable video sequence out of the three was edited and saved as video (.avi) and image (.uis) files. Subsequently, Facialis software
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      (Laboratory for Biomechanics, Swiss Federal Institute of Technology, Zurich, Switzerland) was used to calculate the 3D coordinates of landmarks on the face. FaciShow (Laboratory for Biomechanics, Swiss Federal Institute of Technology, Zurich, Switzerland), a specially designed program,
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      was used to visualise the data processed with the Facialis software. Two-dimensional (2D) and 3D trajectories of each landmark during movement could be presented (Fig. 3) .
      Figure thumbnail gr3
      Fig. 3FaciShow®; (A) visualisation of the distances between landmarks (right brow-central nose and left brow-central nose) during the movement ‘maximal closure of the eyelids and showing of the teeth’ in two dimensions; points represent the movement (mm) of the landmark over time (s); and (B) visualisation of the movement of each landmark during the whole movement ‘maximal closure of the eyelids and showing of the teeth’ in a perspective 3D view.
      All statistical analyses were performed at the Department of Medical Computer Sciences, University of Vienna. The tests were exploratory in nature; thus, no multiple adjustments were performed. Comparisons between two groups were made using Student's t-test after checking the homogeneity of the variances. Differences between more than two groups were assessed using the F-test of a one-way analysis of variance (ANOVA). P≤0.05 was considered statistically significant.

      2. Results

      The mean facial motion in a healthy face ranged from 0.5 to 16.5 mm (Table 3) in the mouth and nose region. Gender and age had statistically significant impacts on facial motions and on the static facial morphology.
      Table 3Mean distances (mm) between the points indicated at rest (D(0)) and during movement for (A) males and females and (B) all age groups (n=24). (A: in each set of three columns, the left-hand column is the data for males and females, the middle column is the data for males only and the right-hand column is the data for females only; B: in each set of three columns, the left-hand column is the data for subjects aged 20–30 years, the middle column is the data for subjects aged 40–50 years and the right-hand column is the data for subjects aged 60–70 years)
      (A)
      Frontal regionD (0)Maximal lifting of the eyebrows
      BP–CN42.643.941.27.88.67.0
      Eye regionD (0)Closure of the eyelids as in sleep
      UE–LE7.67.97.27.68.07.2
      Nose/mouthD (0)Smiling with lips closedMaximal showing of teethSmiling showing teethPursing of the lips
      TR–MC109.0113.3104.89.010.37.615.116.513.712.914.011.710.910.011.8
      CN–AN42.043.739.52.52.62.45.96.94.93.84.53.1
      CN–MU66.470.562.33.43.83.06.88.05.65.35.94.7
      TR–PH134.3139.2129.32.83.12.45.66.05.24.14.63.6
      CN–ML77.180.673.54.65.43.87.78.37.15.66.24.9
      MC–PH36.539.726.97.06.37.7
      (B)
      Frontal regionD (0)Maximal lifting of the eyebrows
      BP–CN43.844.639.37.17.38.9
      Eye regionD (0)Closure of eyelids as in sleep
      UE–LE7.96.78.17.96.78.1
      Nose/mouthD (0)Smiling with lips closedMaximal showing of the teethSmiling showing teethPursing the lips
      TR–MC108.7109.7107.88.88.89.316.013.915.511.312.315.09.612.410.8
      CN–AN41.143.445.92.12.03.54.45.57.72.53.25.6
      CN–MU61.667.070.72.73.14.45.36.88.24.14.37.5
      TR–PH133.4135.0134.52.32.23.75.15.06.73.34.34.6
      CN–ML73.678.079.64.13.76.19.17.06.95.05.06.7
      MC–PH36.036.337.16.27.17.6

      2.1 Gender

      In the resting position, men generally showed statistically significantly larger distances between all landmarks than did women (Table 3(A)). The difference between male and female subjects was, on average, 6.25±2.96 mm (8.26%; P<0.05).
      Larger movements of the face were also observed in males than in females, by, on average, 1.40±0.73 mm (15.08%). A statistically significant difference was seen in the mouth region, where men showed larger movements of the mouth corner than women (tragus–mouth corner: men: 16.5±1.87 mm (14.6%); women: 13.7±3.11 mm (13.1%); P<0.05). Moreover, the excursion of the upper lip was statistically significantly larger in men than in women (central nose–upper lip: men: 7.9±2.55 mm (11.2%); women: 5.6±1.87 mm (8.9%); P<0.05). One exception was observed: on pursing the lips women were found to have a statistically significantly larger excursion than men (tragus point–mouth corner: men: 10.0±3.56 mm (8.8%); women: 11.8±2.16 mm (11.2%); P<0.04; and mouth corner–philturm: men: 6.3±2.39 mm (15.8%); women: 7.7±1.69 mm (28.0%); P<0.04).

      2.2 Age groups

      The effect of age was found to be statistically significant. The mean difference at rest between the oldest age group and subjects in their 20s is 5.68±3.29 mm (6.94%) and the mean difference between the oldest age group and subjects in their 40s is 2.52±1.26 mm (e.g. 4.40%) (Table 3(B)).
      Statistically significant differences between the age groups at rest were seen in the nose region, which was represented by the distance central nose–ala of the nose (20–30 years: 41.1±3.07 mm; 40–50 years: 43.4±2.62 mm; 60–70 years: 45.9±5.29 mm, P<0.05). Moreover, at rest, the oldest age group was observed to have the statistically significantly largest distances in the mouth region, as seen by the distance central nose–upper lip (20–30 years: 61.6±3.91 mm; 40–50 years: 67.0±4.40 mm; 60–70 years: 70.7±8.43 mm; P<0.05). There was one exception in the frontal region: the oldest age group showed statistically significantly shorter distances between the central nose and the brow (20–30 years: 43.8±3.42 mm; 40–50 years: 44.6±2.97 mm; 60–70 years: 39.3±4.91 mm; P<0.05).
      The effect of age on facial movement was also found to be statistically significant. Subjects aged between 60 and 70 years demonstrated the statistically significantly largest movements in the face. They showed, on average, larger facial movements than subjects in their 20s (by 2.10±1.09 mm (33.36%)) and subjects in their 40s (by 1.86±0.60 mm (30.54%)). They were observed to have particularly large changes in the nose region. Statistically significant differences were seen in the distance central nose–ala of the nose (20–30 years: 4.38±1.67 mm (11.03%); 40–50 years: 5.54±2.37 mm (12.52%); 60–70 years: 7.68±3.16 mm (16.76%); P<0.05). Moreover, they were observed to have larger changes in the mouth region, which is represented by the distance central nose–midlateral point of the upper lip, than the other age groups (20–30 years: 5.33±0.98 mm (8.64%); 40–50 years: 6.76±2.89 mm (9.91%); 60–70 years: 8.20±2.48 mm (11.45%); P<0.05).

      2.3 Asymmetry

      Each of the 24 subjects demonstrated differences between the right and left sides of the face. To evaluate side dominance the mean distances on the right and left were compared, and no statistically significant difference was observed. The symmetry in facial motion was stronger than the symmetry in the resting position. In other words, the face in motion is more symmetrical than the face in its resting position.

      2.3.1 Gender

      An asymmetry of 0.61–4.77 mm (0.54–15.67%) was observed in resting position. No statistically significant influence of gender was found in the asymmetry of the resting position (men: 0.82–4.77 mm; women: 0.61–3.77 mm).
      An asymmetry of 0.69–2.68 mm (0.15–6.56%) was observed in facial movements of males and females (men: 0.69–2.50 mm; women: 0.8–2.68 mm) (Table 4(A)) .
      Table 4(A) Asymmetry (mm) at rest (D(0)) and during movement for (A) males and females and (B) all age groups (n=24). (A: in each set of three columns, the left-hand column is the data for males and females, the middle column is the data for males only and the right-hand column is the data for females only; B: in each set of three columns, the left-hand column is the data for subjects aged 20–30 years, the middle column is the data for subjects aged 40–50 years and the right-hand column is the data for subjects aged 60–70 years)
      (A)
      Frontal regionD(0)Maximal lifting of the eyebrows
      BP–CN3.03.42.71.11.21.0
      Eye regionD(0)Closure of eyelids as in sleep
      UE–LE0.80.80.80.00.00.0
      Nose/mouthD(0)Smiling with lips closedMaximal showing of the teethSmiling showing teethPursing the lips
      TR–MC3.84.33.31.92.41.32.12.12.12.62.52.72.12.02.2
      CN–AN1.81.62.00.90.81.01.41.51.31.11.11.1
      CN–MU1.21.31.10.80.80.81.41.41.31.01.01.0
      TR–PH3.84.33.20.91.00.81.21.01.40.90.90.9
      CN–ML1.01.30.71.21.21.21.31.31.31.00.71.2
      MC–PH2.12.02.21.20.91.4
      (B)
      Frontal regionD(0)Maximal lifting of the eyebrows
      BP–CN2.74.22.10.71.11.4
      Eye regionD(0)Closure of eyelids as in sleep
      UE–LE1.00.70.70.00.00.0
      Nose/mouthD(0)Smiling with lips closedMaximal showing of the teethSmiling showing teethPursing the lips
      TR–MC3.25.13.22.41.41.82.12.41.62.81.93.11.43.21.7
      CN–AN1.01.92.50.70.91.11.21.01.90.81.11.4
      CN–MU0.91.21.50.70.61.00.91.41.71.01.10.9
      TR–PH3.73.73.90.90.80.71.50.71.41.10.80.8
      CN–ML1.00.91.20.90.80.90.61.71.51.10.81.0
      MC–PH2.02.22.00.91.61.0

      2.3.2 Age groups

      Among the age groups the asymmetry between the right and left sides of the face ranged from 0.71 to 4.66 mm for subjects aged 20–30 years, 0.68–5.62 mm for subjects aged 40–50 years and 0.65–4.15 mm for subjects aged 60–70 years. No statistically significant differences between the right and left sides were found (Table 4(B)).
      Age has no statistically significant impact on asymmetry during facial motion: asymmetry in those aged 20–30 years ranged from 0.52 to 2.78 mm; in those aged 40–50 years it ranged from 0.65 to 3.2 mm; and in those aged 60–70 years it ranged from 0.65 to 3.10 mm (Table 4(B)).

      3. Discussion

      3.1 Population sample

      It is reported that the three major geographical races—African, Asian and European—have each evolved a characteristic set of facial features,
      • Goetz P.W.
      Face.
      and this has been confirmed by anthropometric research.
      • Farkas L.G.
      • Farkas L.G.
      Accuracy of anthropometric measurements: past, present, and future.
      • Farkas L.G.
      • Ngim R.C.
      • Lee S.T.
      The fourth dimension of the face: a preliminary report of growth potential in the face of the Chinese population of Singapore.
      One ethnic group is predominant in a number of European countries (e.g. France, Germany, Italy, Spain, Scandinavia, the Czech Republic, Hungary and Poland), in comparison with the multi-ethnic population of North America. Therefore, to create a representative population sample, we studied healthy facial movements in subjects only of Middle European descent, and individuals living in the eastern part of Austria were invited to take part in this study.

      3.2 3D system

      Standards already exist in the anthropometric measurement of faces. This study provides the first standards for the 3D assessment of healthy facial movements. This objective 3D video-analysis technique
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      used a mirror system, a calibration grid, a commercial digital video camera, analysis software (Facialis) and visualisation software (FaciShow) to capture all facial dimensions and movements of interest.
      Our method of assessing movements of dynamic landmarks on the face differs from those used in recent research.
      • Bajaj-Luthra A.
      • Mueller T.
      • Johnson P.C.
      Quantitative analysis of facial motion components: anatomic and nonanatomic motion in normal persons and in patients with complete facial paralysis.
      • Johnson P.J.
      • Bajaj-Luthra A.
      • Llull R.
      • et al.
      Quantitative facial motion analysis after functional free muscle reanimation procedures.
      Those studies used 2D measurements, an approach that is nowadays considered unacceptable. According to Frey et al., important information is lost and accuracy of measurement is sacrificed by projecting 3D data into 2D.
      • Frey M.
      • Jenny A.
      • Giovanoli P.
      • et al.
      Development of a new documentation system for facial movements as a basis for the international registry for neuromuscular reconstruction in the face.
      Moreover, Frey et al. stated that only a 3D analysis could adequately and objectively document the extremely complex facial movements.
      • Frey M.
      • Jenny A.
      • Giovanoli P.
      • et al.
      Development of a new documentation system for facial movements as a basis for the international registry for neuromuscular reconstruction in the face.
      Gross et al. found significant differences between the 3D and 2D amplitudes of measurements of, on average, up to 3.0 mm when facial motion was greatest.
      • Gross M.M.
      • Trotman C.A.
      • Moffatt K.S.
      A comparison of three-dimensional and two-dimensional analyses of facial motion.
      Further, they also pointed out that projection error arose from the inability of the 2D method to detect any anterior–posterior motion of the landmarks. Projection errors also resulted from misalignment between the frontal plane of the face and the image plane of the camera sensors.
      These errors might have affected Burres' study, in which a 2D technique was applied to 30 normal subjects.
      • Burres S.A.
      Facial biomechanics: the standards of normal.
      He found that the maximal movement of any specific point on the face did not exceed 10 mm. In contrast, when we assessed facial movements with 3D video-analysis, we found that the maximal excursion of a landmark (the mouth corner) was 27.97 mm, in male subjects aged between 60–70 years, when smiling showing teeth.
      2D error was also found in photogrammetry,
      • Farkas L.G.
      • Bryson W.
      • Klotz J.
      Is photogrammetry of the face reliable?.
      a widely used technique in which measurements of faces are taken from standardised photographs. The photographs were not sharp enough to allow accurate identification and marking of some landmarks on the photographs in all cases (e.g. ala of the nose and upper and lower eyelids). Therefore, in order to produce the best results, the subject had to be photographed with the landmarks marked on the skin.
      • Farkas L.G.
      • Bryson W.
      • Klotz J.
      Is photogrammetry of the face reliable?.
      The marking of landmarks in our 3D video-analysis was done before videotaping the subjects' facial movements. This procedure was fully accepted by the participants, who were cooperative and informed. Facial movements were not influenced by the marking method used in this study.
      Many systems have used markers up to 6.4 mm in diameter to determine landmarks on the face (Kohn et al.
      • Kohn L.A.
      • Cheverud J.M.
      • Bhatia G.
      • et al.
      Anthropometric optical surface imaging system repeatability, precision, and validation.
      6.4 mm; Bajaj-Luthra et al.
      • Bajaj-Luthra A.
      • Mueller T.
      • Johnson P.C.
      Quantitative analysis of facial motion components: anatomic and nonanatomic motion in normal persons and in patients with complete facial paralysis.
      5.0 mm; Johnson et al.
      • Johnson P.J.
      • Bajaj-Luthra A.
      • Llull R.
      • et al.
      Quantitative facial motion analysis after functional free muscle reanimation procedures.
      5.0 mm; Gross et al.
      • Gross M.M.
      • Trotman C.A.
      • Moffatt K.S.
      A comparison of three-dimensional and two-dimensional analyses of facial motion.
      5.0 mm; and Trotman et al.
      • Trotman C.A.
      • Gross M.M.
      • Moffatt K.
      Reliability of a three-dimensional method for measuring facial animation: a case report.
      4.0 mm). In this study a permanent marker was used to place a 2 mm black dot at each of 16 dynamic landmarks, which proved to be very practical. The dots did not interfere with facial motion at any time during the measurement and were definitely more advantageous than the landmarks (plastic balls) previously used to mark dynamic points. The three static points were essential for calibrating the space in the mirror system and were marked with plastic light balls of 5.0 mm in diameter. The large size of the static landmarks has no effect on measurement bias, since the centre-marking principle was used to determine their positions.

      3.3 Standardised procedures

      In order to maximally eliminate errors from all aspects, the procedures in this study were standardised. All recordings were made in the same room, in the same chair, by the same examiner, at the same time of day—between 14:00 and 16:00—and under the same lighting conditions. Reproducible and standardised landmarks,
      • Frey M.
      • Giovanoli P.
      • Gerber H.
      • et al.
      Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
      set up and standardised in the International Registry for Neuromuscular Reconstruction in the Face,
      • Frey M.
      • Jenny A.
      • Giovanoli P.
      • et al.
      Development of a new documentation system for facial movements as a basis for the international registry for neuromuscular reconstruction in the face.
      were selected and marked. The 3D video-analysis is able to visualise the exact coordinates of each point in the mirror complex. Moreover, it is also able to illustrate the velocity and direction of each landmark within the 3D space of interest.
      This objective method differs from those of previous studies that were able to analyse differences between the resting and maximal positions only, except for Trotman et al.,
      • Trotman C.A.
      • Gross M.M.
      • Moffatt K.
      Reliability of a three-dimensional method for measuring facial animation: a case report.
      who demonstrated a video-analysis system, which provided data on facial movements in 3D. However, in their system, absolute measurements of changes in distance between landmarks were impossible.
      Another advantage of this video system compared with a laser system is the real-time taping of the subjects. Vannier et al. reported that the time a person can remain motionless and expressionless is approximately 1 s.
      • Vannier M.W.
      • Pilgram T.K.
      • Bhatia G.
      • et al.
      Facial surface scanner.
      Thus, using a laser scanning system, which usually needs around 8–30 s on average,
      • Yamada T.
      • Sugahara T.
      • Mori Y.
      • et al.
      Rapid three-dimensional measuring system for facial surface structure.
      might cause motion artefacts. The real-time recording used by this 3D video system prevents motion artefacts and fatigue of the subject's mimic system.
      This sensitive standardised 3D video-analysis system has been used to assess the functional recovery of the facial nerve. It has been applied in clinical practice at the Division of Plastic and Reconstructive Surgery, Department of Surgery, General Hospital Vienna since 1998. Degrees of asymmetry have been successfully quantified and evaluated before and after surgery in patients suffering from facial paralysis.

      3.4 Mean facial movements

      Generally, gender and age seem to have a great impact on facial movements. Men and older subjects demonstrated larger distances both at rest and in facial movements.

      3.4.1 Gender

      Men have statistically significantly larger facial distances than women. This might be due to increased growth of the male facial musculature and skull as compared with the lighter features of the female face, which has been reported to be about four-fifths the size of the male face.
      • Farkas L.G.
      • McNeill D.
      In this study, the distance between the right and left tragus points, representing the width of the face, shows a noticeable difference between males and females (men: 162.2±4.76 mm; women: 149.4±4.76 mm).
      Farkas observed that the eye-fissure distance (upper eyelid–lower eyelid) in females is 0.9 mm larger than in males.
      • Farkas L.G.
      However, in this study, differences in the eye-fissure distance between the sexes were not statistically significant: women had a mean of 7.22 mm and men had a mean of 7.93 mm.
      The shortening of the distance between the tragus point and the mouth corner when smiling showing teeth is, on average, 11.7 mm for women and 14.0 mm for men. Women show an excursion of only 83.5% of that shown by men. The larger excursion of the mouth corner in men probably results from their larger facial dimensions. Paletz et al. demonstrated similar results using a 2D video system.
      • Paletz J.L.
      • Manktelow R.T.
      • Chaban R.
      The shape of a normal smile: implications for facial paralysis reconstruction.
      They reported that the movement when smiling showing teeth in females was 90% of that in males, measuring absolute changes only.

      3.4.2 Age

      At rest and in motion, most distances are statistically significantly larger (by up to 9.28 mm) in subjects aged between 60 and 70 years than in younger subjects. This finding might relate to changes occurring in the face with age: the nasolabial folds become more pronounced, the nose gets bigger with a down-turned nasal tip and the skin becomes thinner and more wrinkled.
      • Medina J.J.
      Age-related changes in the face suggest that faces with wrinkled and surplus skin will show larger excursions of landmarks during facial movements, which is assumed to be due to the long-term effects of gravity.
      Statistically significant differences between the age groups can be clearly seen in excursions around the nose and mouth, particularly in the distances between the central nose and the mouth (upper lip and lower lip): whereas subjects aged 60–70 years show, on average, the largest excursions of 12.12%, subjects aged 40–50 years show average excursions of 8.23% and subjects aged 20–30 years show average excursions of 7.31%.

      3.5 Asymmetry

      Differences between the right and left sides of the face are observed in all facial motions. Gender and age have no statistically significant impact on the asymmetry of facial dimensions. A slight left dominance is seen in facial movements in both sexes and all age groups. Slight left dominance in resting position is also observed in both sexes and in the first age group (20–30 years); however a right dominance in resting position increases with increasing age (40–50 and 60–70 years). Often facial asymmetry cannot be seen with the human eye, but marginal differences between the right and left sides of the face of up to 4.77 mm at rest and up to 2.68 mm during facial motion can be measured.
      One of the explanations for facial asymmetry was suggested by Peck et al.
      • Peck S.
      • Peck L.
      • Kataja M.
      Skeletal asymmetry in esthetically pleasing faces.
      They found that the strongest asymmetry in well-balanced faces is in the distance between the tragus and a point at the angulus maxillae. They suggested that, since these two landmarks are related anatomically through dental occlusion, tooth position might have a significant impact on the asymmetry of the face. Melnik suggested that craniofacial asymmetry might be due to relative growth imbalances between the right and left sides.
      • Melnik A.K.
      A cephalometric study of mandibular asymmetry in a longitudinally followed sample of growing children.
      Asymmetry seems to be the rule rather than the exception
      • Williamson E.H.
      • Simmons M.D.
      Mandibular asymmetry and its relation to pain dysfunction.
      and an intrinsic characteristic of the human face.
      • Lu K.H.
      Harmonic analysis of the human face.
      Our finding that asymmetry at rest may be up to 15.67% does not match the findings of Lu, who claimed that only facial asymmetry greater than 3% is clinically discernable.
      • Lu K.H.
      Harmonic analysis of the human face.
      The discrepancy between this and Lu's study might be due to different systems of measuring the facial dimensions. 2D errors may have been important in Lu's study, where radiographs of heads were used to measure facial dimensions.
      Right facial dominance was reported in asymmetry analyses by Shah and Joshi
      • Shah S.M.
      • Joshi M.R.
      An assessment of asymmetry in the normal craniofacial complex.
      and Ferrario et al.
      • Ferrario V.F.
      • Sforza C.
      • Poggio C.E.
      • et al.
      Distance from symmetry: a three-dimensional evaluation of facial asymmetry.
      Woo suggested that right craniofacial dominance may be naturally favoured for neuroanatomical development reasons.
      • Woo T.L.
      On the asymmetry of the human skull.
      However, other studies have found a tendency toward left-side dominance, which was interpreted to imply a genetic predisposition.
      • Williamson E.H.
      • Simmons M.D.
      Mandibular asymmetry and its relation to pain dysfunction.
      • Vig P.S.
      • Hewitt A.B.
      Asymmetry of the human facial skeleton.
      • Chebib F.S.
      • Chamma A.M.
      Indices of craniofacial asymmetry.
      Asymmetry in facial motion, of up to 2.68 mm, was a common finding in this study. However, statistically significant dynamic asymmetry was not observed.
      Dynamic asymmetry was also found by Scriba et al., who reported an asymmetry of 7–9% in motion, which they considered to be physiological; like us, they found no side dominance.
      • Scriba H.
      • Stoeckli S.J.
      • Veraguth D.
      • et al.
      Objective evaluation of normal facial function.
      Asymmetry in movements was also observed by Ferrario et al. and was particularly evident in the middle and lower parts of the face.
      • Ferrario V.F.
      • Sforza C.
      • Poggio C.E.
      • et al.
      Distance from symmetry: a three-dimensional evaluation of facial asymmetry.
      However, owing to differences in methods, measurements and sample characteristics (age, sex, race), it is difficult to compare these studies. It may be that some of the differences are methodological in nature.

      4. Conclusion

      In this study asymmetry was found in healthy subjects, which shows that normal faces have natural asymmetry.
      The evaluation of facial movements in 24 healthy volunteers showed that males have greater facial motion than females. Moreover, this study demonstrated that age has an impact on facial dimensions, since the subjects aged between 60 and 70 years had statistically significantly larger distances between landmarks than the other age groups.
      Ethnicities seem to have a significant impact on static and dynamic facial dimensions. Subsequent studies will find healthy standard values for other ethnic populations using this 3D video-analysis system.
      Furthermore, the application of this objective 3D video-analysis system to provide a standardised outcome measure of surgical success is the focus of present and future research.

      Acknowledgements

      This study was conducted at the Division of Plastic and Reconstructive Surgery, Department of Surgery, General Hospital Vienna, Medical School, University of Vienna, Austria, in collaboration with the Laboratory for Biomechanics of the Swiss Federal Institute of Technology, Zurich, Switzerland. The study was supported by grant #GZ308.974, entitled ‘3-Dimensional analysis of movements in reconstructive facial surgery’, from the Austrian Federal Ministry of Science and Transport, Vienna, Austria. This project was designed to establish an international reference centre for neuromuscular reconstructions in the face and to build up an international development project and a prospective multicentre study. The study was supported in part by Grant #90296554 (doctoral dissertation grant for Chieh-Han John Tzou, Förderungsstipendium an der Medizinischen Fakultät der Universität Wien 1999) from the Dean of the Medical School, University of Vienna.

      References

        • Takasaki H.
        Moire topography.
        Appl Optics. 1970; 9: 1467-1472
        • Burke P.H.
        • Banks P.
        • Beard L.F.
        • et al.
        Stereophotographic measurement of change in facial soft tissue morphology following surgery.
        Br J Oral Surg. 1983; 21: 237-245
        • Deacon A.T.
        • Anthony A.G.
        • Bhatia S.N.
        • et al.
        Evaluation of a CCD-based facial measurement system.
        Med Inform (Lond). 1991; 16: 213-228
        • Neely J.G.
        • Cheung J.Y.
        • Wood M.
        • et al.
        Computerized quantitative dynamic analysis of facial motion in the paralyzed and synkinetic face.
        Am J Otol. 1992; 13: 97-107
        • Sato K.
        • Inokuchi I.
        A range-imaging system utilizing a nematic liquid crystal mask.
        IEEE 1st ICCV. IEEE, London1987
        • Paletz J.L.
        • Manktelow R.T.
        • Chaban R.
        The shape of a normal smile: implications for facial paralysis reconstruction.
        Plast Reconstr Surg. 1994; 93: 784-789
        • Bush K.
        • Antonyshyn O.
        Three-dimensional facial anthropometry using a laser surface scanner: validation of the technique.
        Plast Reconstr Surg. 1996; 98: 226-235
        • Kohn L.A.
        • Cheverud J.M.
        • Bhatia G.
        • et al.
        Anthropometric optical surface imaging system repeatability, precision, and validation.
        Ann Plast Surg. 1995; 34: 362-371
        • Ras F.
        • Habets L.L.
        • van Ginkel F.C.
        • et al.
        Quantification of facial morphology using stereophotogrammetry—demonstration of a new concept.
        J Dent. 1996; 24: 369-374
        • Bajaj-Luthra A.
        • Mueller T.
        • Johnson P.C.
        Quantitative analysis of facial motion components: anatomic and nonanatomic motion in normal persons and in patients with complete facial paralysis.
        Plast Reconstr Surg. 1997; 99: 1894-1902
        • Bajaj-Luthra A.
        • Van Swearingen J.
        • Thornton R.H.
        • et al.
        Quantitation of patterns of facial movement in patients with ocular to oral synkinesis.
        Plast Reconstr Surg. 1998; 101: 1473-1480
        • Meier-Gallati V.
        • Scriba H.
        • Fisch U.
        Objective scaling of facial nerve function based on area analysis (OSCAR).
        Otolaryngol Head Neck Surg. 1998; 118: 545-550
        • Caruso A.J.
        • Stanhope S.J.
        • McGuire D.A.
        A new technique for acquiring three-dimensional orofacial nonspeech movements.
        Dysphagia. 1989; 4: 127-132
        • Vannier M.W.
        • Pilgram T.K.
        • Bhatia G.
        • et al.
        Quantitative three-dimensional assessment of face-lift with an optical facial surface scanner.
        Ann Plast Surg. 1993; 30: 204-211
        • Ferrario V.F.
        • Sforza C.
        • Poggio C.E.
        • et al.
        Distance from symmetry: a three-dimensional evaluation of facial asymmetry.
        J Oral Maxillofac Surg. 1994; 52: 1126-1132
        • Trotman C.A.
        • Gross M.M.
        • Moffatt K.
        Reliability of a three-dimensional method for measuring facial animation: a case report.
        Angle Orthod. 1996; 66: 195-198
        • Trotman C.A.
        • Faraway J.J.
        • Silvester K.T.
        • et al.
        Sensitivity of a method for the analysis of facial mobility. I. Vector of displacement.
        Cleft Palate Craniofac J. 1998; 35: 132-141
        • Frey M.
        • Jenny A.
        • Giovanoli P.
        • et al.
        Development of a new documentation system for facial movements as a basis for the international registry for neuromuscular reconstruction in the face.
        Plast Reconstr Surg. 1994; 93: 1334-1349
        • Frey M.
        • Giovanoli P.
        • Gerber H.
        • et al.
        Three-dimensional video analysis of facial movements: a new method to assess the quantity and quality of the smile.
        Plast Reconstr Surg. 1999; 104: 2032-2039
        • Goetz P.W.
        Face.
        Encyclopaedia Britannica. 15th. Encyclopaedia Britannica, Chicago1991
        • Farkas L.G.
        Anthropometry of the head and face. Raven Press, New York1994
        • Farkas L.G.
        Accuracy of anthropometric measurements: past, present, and future.
        Cleft Palate Craniofac J. 1996; 33: 10-18
        • Farkas L.G.
        • Ngim R.C.
        • Lee S.T.
        The fourth dimension of the face: a preliminary report of growth potential in the face of the Chinese population of Singapore.
        Ann Acad Med Singapore. 1988; 17: 319-327
        • Johnson P.J.
        • Bajaj-Luthra A.
        • Llull R.
        • et al.
        Quantitative facial motion analysis after functional free muscle reanimation procedures.
        Plast Reconstr Surg. 1997; 100: 1710-1719
        • Gross M.M.
        • Trotman C.A.
        • Moffatt K.S.
        A comparison of three-dimensional and two-dimensional analyses of facial motion.
        Angle Orthod. 1996; 66: 189-194
        • Burres S.A.
        Facial biomechanics: the standards of normal.
        Laryngoscope. 1985; 95: 708-714
        • Farkas L.G.
        • Bryson W.
        • Klotz J.
        Is photogrammetry of the face reliable?.
        Plast Reconstr Surg. 1980; 66: 346-355
        • Vannier M.W.
        • Pilgram T.K.
        • Bhatia G.
        • et al.
        Facial surface scanner.
        IEEE Comput Graph Appl. 1991; : 72-80
        • Yamada T.
        • Sugahara T.
        • Mori Y.
        • et al.
        Rapid three-dimensional measuring system for facial surface structure.
        Plast Reconstr Surg. 1998; 102: 2108-2113
        • McNeill D.
        The Face. Hamish Hamilton Ltd, London1998
        • Medina J.J.
        The clock of ages: Why we age—how we age—winding back the clock. Cambridge University Press, New York1996
        • Peck S.
        • Peck L.
        • Kataja M.
        Skeletal asymmetry in esthetically pleasing faces.
        Angle Orthod. 1991; 61: 43-48
        • Melnik A.K.
        A cephalometric study of mandibular asymmetry in a longitudinally followed sample of growing children.
        Am J Orthod Dentofac Orthop. 1992; 101: 355-366
        • Williamson E.H.
        • Simmons M.D.
        Mandibular asymmetry and its relation to pain dysfunction.
        Am J Orthod. 1979; 76: 612-617
        • Lu K.H.
        Harmonic analysis of the human face.
        Biometrics. 1965; 21: 491-505
        • Shah S.M.
        • Joshi M.R.
        An assessment of asymmetry in the normal craniofacial complex.
        Angle Orthod. 1978; 48: 141-148
        • Woo T.L.
        On the asymmetry of the human skull.
        Biometrika. 1931; 22: 324-352
        • Vig P.S.
        • Hewitt A.B.
        Asymmetry of the human facial skeleton.
        Angle Orthod. 1975; 45: 125-129
        • Chebib F.S.
        • Chamma A.M.
        Indices of craniofacial asymmetry.
        Angle Orthod. 1981; 51: 214-216
        • Scriba H.
        • Stoeckli S.J.
        • Veraguth D.
        • et al.
        Objective evaluation of normal facial function.
        Ann Otol Rhinol Laryngol. 1999; 108: 641-644